National-instruments 24-Bit Half/Full-Bridge Analog Input Module NI 923 Instrukcja Użytkownika Strona 22

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 36
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 21
NI 9237 Operating Instructions and Specifications 22 ni.com
Type of ADC..................................... Delta-Sigma (with analog
prefiltering)
Sampling mode ................................. Simultaneous
Internal master timebase (f
M
)
Frequency ................................... 12.8 MHz
Accuracy..................................... ±100 ppm max
Data rate range (f
s
) using internal master timebase
Minimum.................................... 1.613 kS/s
Maximum ................................... 50 kS/s
Data rate range (f
s
) using external master timebase
Minimum.................................... 390.625 S/s
Maximum ................................... 51.3 kS/s
Data rates
1
(f
s
)................................... , n = 1, 2, …, 31
Typical input range ........................... ±25 mV/V
Scaling coefficient ............................ 2.9802 nV/V per LSB
1
The data rate must remain within the appropriate data rate range. Refer to the
Understanding NI 9237 Data Rates section for more information.
f
M
256÷
n
---------------------
Przeglądanie stron 21
1 2 ... 17 18 19 20 21 22 23 24 25 26 27 ... 35 36

Komentarze do niniejszej Instrukcji

Brak uwag