© National Instruments Corporation 7 SCXI-1125 Calibration Procedure
calOp
ND_SELF_CALIBRATE
setOfCalConstants
ND_USER_EEPROM_AREA
calRefVolts0.0
4. Call
AI_Configure
to put the E Series device in bipolar mode. Set the
following parameters:
deviceNumberThe value assigned by MAX
chan0
inputMode0
inputRange0
polarity0
driveAIS0
5. Call
MIO_Config
to enable dithering on the E Series device. Set the
following parameters:
deviceNumberThe value assigned by MAX
dither1
useAMUX0
6. Call
SCXI_Single_Chan_Setup
to configure the module for
single-channel measurements. Set the following parameters:
SCXIchassisIDThe value assigned by MAX
moduleSlot1
moduleChanThe channel you want to adjust
DAQdeviceNumberThe device number assigned by MAX for
the E Series DAQ device
7. Call
SCXI_Set_Gain
.
Set the following parameters:
SCXIchassisIDThe ID of the SCXI chassis obtained when you
configure the chassis with MAX
moduleSlot1, because the SCXI-1125 should be installed in
slot 1
channelThe same channel number from step 6
gainThe gain value you want to adjust. Refer to Table 1 for
acceptable gain values
8. Call
SCXI_Configure_Filter
to configure the filter setting on the
module. Set the following parameters:
SCXIchassisIDThe ID of the SCXI chassis obtained when you
configure the chassis with MAX
moduleSlot1
channelThe same channel number from step 7