Appendix A Specifications
© National Instruments Corporation A-3 SCXI-1125 User Manual
Absolute Accuracy
Nominal Range
Overall
Gain
% of Reading
Offset
System Noise (peak, 3 sigma)
Temperature Drift
Absolute
Accuracy at
Full Scale
(mV)
1
Single Point Average
Typical Max 4 Hz 10 kHz 4 Hz 10 kHz
% of
Reading/°C
Offset
(uV/°C)
±1000 VDC
1, 2, 3
0.005 0.3996 1.25 854 mV 115 mV 1.62 V 25 mV 401 mV 0.0034 132 mV 14.0V
±250 V
4
0.02 0.2548 0.65 250 mV 309 mV 478 mV 7 mV 100 mV 0.0029 44 mV 2.0V
±100 V
4
0.05 0.2548 0.65 100 mV 12 mV 183 mV 2.6 mV 41 mV 0.0029 22 mV 790mV
±50 V
4
0.1 0.2548 0.65 50 mV 6 mV 111 mV 1.3 mV 21 mV 0.0029 11 mV 400mV
±25 V
4
0.2 0.2548 0.65 25 mV 3 mV 48 mV 641 μV 11 mV 0.0029 4.4 mV 200mV
±10 V
4
0.5 0.2548 0.65 10 mV 1 mV 19.1 mV 238 μV 4.1 mV 0.0029 2.2 mV 80.0mV
±5 V 1 0.07 0.52 5.0 mV 530 μV 8.6 mV 122 μV 2.1 mV 0.0027 251 μV 33.0mV
±2.5 V 2 0.07 0.52 2.5 mV 254 μV 4.3 mV 60 μV 1.1 mV 0.0027 126 μV 16.5mV
±1 V 5 0.07 0.52 1.0 mV 110 μV 1.7 mV 24 μV 403 μV 0.0027 51 μV 6.6mV
±500 mV 10 0.07 0.52 500 μV 69 μV 882 μV 13 μV 202 μV 0.0027 26 μV 3.31mV
±250 mV 20 0.07 0.52 250 μV 32.0 μV 474 μV 6.3 μV 101 μV 0.0027 13 μV 1.65mV
±100 mV 50 0.07 0.52 110 μV 11 μV 180 μV 2.4 μV 41 μV 0.0027 5.5 μV 670μV
±50 mV 100 0.07 0.52 60 μV 6.2 μV 88.2 μV 1.3 μV 21 μV 0.0027 3.0 μV 340μV
±25 mV 200 0.08 0.52 30 μV 2.6 μV 47.9 μV 0.60 μV 11 μV 0.0027 1.7 μV 170μV
±20 mV 250 0.08 0.53 25 μV 2.3 μV 37.1 μV 0.50 μV 7 μV 0.0027 1.5 μV 140μV